RecallRadar

FDA Device recall Z-2054-2013

Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015…

On 2013-09-04 the FDA classified a Class II recall of Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015… by Stryker Craniomaxillofacial Division, citing stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadver….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2054-2013
ClassificationClass II
Statusterminated
Initiated2013-07-30
Published2013-09-04
Last updated2026-09-13 11:53 UTC
CompanyStryker Craniomaxillofacial Division
DistributionUS

Product

Stryker Part Numbers 79-43903 (3cc), 79-43905 (5cc), 79-43910 (10 cc), 79-43915 (15cc), 397003 (3cc), 397005 (5cc), 397010 (10 cc), 397015 (15 cc) HydroSet Injectable HA Bone Substitute Sterile R/EO, nonpyrogenic, Product of Ireland Stryker Injectable Cement is a self-setting, calcium phosphate cement intended for use in the repair of neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects as well as in the augmentation or restoration of bony contour in the craniofacial skeleton. The powder is mixed with the sodium phosphate/PVP solution provided in the kit and loaded into the empty cement delivery syringe to be applied directly to the site for void filling or augmentation.

Reason

Stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadvertent implantation of silica gel granulate which could cause foreign body reaction including inflammation and might lead to the need for revision surgery.

Identifiers

code_info
Part 79-43903, Lot: IC01116, IC01130, IC01204, IC01233, IC01238 Part 79-43905 Lot: IC01084, IC01085, IC01086, IC01093,…Part 79-43903, Lot: IC01116, IC01130, IC01204, IC01233, IC01238 Part 79-43905 Lot: IC01084, IC01085, IC01086, IC01093, IC01123, IC01150, IC01166, IC01167, IC01168, IC01180, IC01188, IC01189, IC01190, IC01198, IC01199, IC01200, IC01213, IC01214, IC01229, IC01230, IC01239, IC01240, IC01245, IC01246, IC01270 Part 79-43910 Lot: IC01087, IC01090, IC01091, IC01153, IC01154, IC01163, IC01164, IC01165, IC01179, IC01183, IC01184, IC01201, IC01202, IC01203, IC01231, IC01243, IC01244, IC01248, IC01249, IC01261, IC01264, Part 79-43015 Lot: IC01100, IC01114, IC01161, IC01162, IC01182, IC01211, IC01212, IC01227, IC01228, IC01241, IC01242, IC01252, IC01253, IC01254, IC01267 Part 397003 Lot: IC01092, IC01101, IC01102, IC01103, IC01104, IC01105, IC01106, IC01107, IC01115, IC01127, IC01131, IC01205, IC01220, IC01232, IC01234 Part 397005 Lot: IC01088, IC01094, IC01095, IC01096, IC01097, IC01098, IC01110, IC01111, IC01118, IC01119, IC01120, IC01132, IC01133, IC01134, IC01135, IC01136, IC

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2054-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.