FDA Device recall Z-2052-2012
Sarns Modular Perfusion System 9000, 100V with color screen
- FDA Device
- Class II
- terminated
- Published 2012-07-25
On 2012-07-25 the FDA classified a Class II recall of Sarns Modular Perfusion System 9000, 100V with color screen by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malf….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2052-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-18 |
| Published | 2012-07-25 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Sarns Modular Perfusion System 9000, 100V with color screen. The Sarns Perfusion System 9000 is indicated for use as a heart-lung system for pumping blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Reason
Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over sensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syst
Identifiers
| code_info | Catalog Number 164280 and serial number 1019-5379 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2052-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.