FDA Device recall Z-2048-2016
EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080
- FDA Device
- Class II
- terminated
- Published 2016-06-29
On 2016-06-29 the FDA classified a Class II recall of EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080 by Ev3, citing medtronic is issuing a medical device recall to return product potentially associated with an outer box mislabel issue for two Protege EverFiex" Self-expanding Peripheral Stent Sy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2048-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-05-23 |
| Published | 2016-06-29 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ev3 |
| Distribution | US |
Product
EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a self-expanding Nitinol stent system intended for permanent implantation. The self-expanding stent is made of a nickel titanium alloy (Nitinol) and comes pre-mounted on a 6 F, 0.035" over-the-wire delivery system. The stent is cut from a Nitinol tube in an open lattice design, and has tantalum radiopaque markers at the proximal and distal ends of the stent. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency
Reason
Medtronic is issuing a medical device recall to return product potentially associated with an outer box mislabel issue for two Protege EverFiex" Self-expanding Peripheral Stent Systems, model PRB35-08-040- 080, from lot A164580. The size on the outer box is labeled as 8x40x80mm while potentially containing a 6x40x120mm device as indicated by both the inner pouch and device labels. This potential issue is limited to two EverFiex stent system devices from lot A164580 that were repackaged by Medtronic and distributed. This does not affect any other Medtronic product or implantable devices.
Identifiers
| code_info | lot: A164580 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2048-2016%22