RecallRadar

FDA Device recall Z-2040-2023

EliA GliadinDP IgG Well in the Phadia 2500EE instrument (Material Number 12410002) EliA GliadinDP IgG is intended for the in vitro semi-qua…

On 2023-07-05 the FDA classified a Class II recall of EliA GliadinDP IgG Well in the Phadia 2500EE instrument (Material Number 12410002) EliA GliadinDP IgG is intended for the in vitro semi-qua… by Phadia Us, citing higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2040-2023
ClassificationClass II
Statusongoing
Initiated2023-05-01
Published2023-07-05
Last updated2026-09-13 11:56 UTC
CompanyPhadia Us
DistributionCA, MI, NJ

Product

EliA GliadinDP IgG Well in the Phadia 2500EE instrument (Material Number 12410002) EliA GliadinDP IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease Material Number 14553901

Reason

Higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument

Identifiers

code_infoUDI: 07333066010847 All lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2040-2023%22