FDA Device recall Z-2036-2014
Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended f…
- FDA Device
- Class III
- terminated
- Published 2014-07-16
On 2014-07-16 the FDA classified a Class III recall of Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended f… by Steris, citing STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the d….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2036-2014 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2014-06-30 |
| Published | 2014-07-16 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Steris |
| Distribution | US |
Product
Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.
Reason
STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.
Identifiers
| code_info | Serial Number Range: 3632511020 - 3613214010 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.