FDA Device recall Z-2036-2012

Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm…

On 2012-07-25 the FDA classified a Class II recall of Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm… by Teleflex Medical, citing speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2036-2012
ClassificationClass II
Statusterminated
Initiated2012-06-12
Published2012-07-25
Last updated2026-09-13 11:53 UTC
CompanyTeleflex Medical
DistributionUS

Product

Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.

Reason

Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.

Identifiers

code_infoCatalog #425-00, 425-10 and 425-30.