FDA Device recall Z-2036-2012
Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm…
- FDA Device
- Class II
- terminated
- Published 2012-07-25
On 2012-07-25 the FDA classified a Class II recall of Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm… by Teleflex Medical, citing speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2036-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-12 |
| Published | 2012-07-25 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Teleflex Medical |
| Distribution | US |
Product
Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.
Reason
Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
Identifiers
| code_info | Catalog #425-00, 425-10 and 425-30. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.