FDA Device recall Z-2034-2017

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

On 2017-05-10 the FDA classified a Class II recall of Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan by Sysmex America, citing reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reag….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2034-2017
ClassificationClass II
Statusterminated
Initiated2017-03-23
Published2017-05-10
Last updated2026-09-13 11:54 UTC
CompanySysmex America
DistributionUS

Product

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

Reason

Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.

Identifiers

code_info
**Model: XN-10, Product Number: AP795756, Serial Numbers: 11001  29656, 29668-29786, 29788-29798, 29800-29801, 29803,…**Model: XN-10, Product Number: AP795756, Serial Numbers: 11001  29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171