RecallRadar

FDA Device recall Z-2032-2025

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Pro…

On 2025-07-16 the FDA classified a Class I recall of Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Pro… by Baxter Healthcare, citing the Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury fr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2032-2025
ClassificationClass I
Statusongoing
Initiated2025-05-30
Published2025-07-16
Last updated2026-09-13 11:57 UTC
CompanyBaxter Healthcare
Distributionn/a

Product

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Reason

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Identifiers

code_infoAll LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2032-2025%22