FDA Device recall Z-2032-2013
00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9, Rx, Sterile
- FDA Device
- Class II
- terminated
- Published 2013-08-28
On 2013-08-28 the FDA classified a Class II recall of 00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9, Rx, Sterile by Zimmer, citing surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2032-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-07 |
| Published | 2013-08-28 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9, Rx, Sterile; 00597204110 Articular Surface CR ART SURF 56/STRIPE GRN 10, Rx, Sterile; 00597204112 Articular Surface CR ART SURF 56/STRIPE GRN 12, Rx, Sterile; 00597204114 Articular Surface CR ART SURF 56/STRIPE GRN 14, Rx, Sterile; 00597204117 Articular Surface CR ART SURF 56/STRIPE GRN 17, Rx, Sterile; 00597204120 Articular Surface CR ART SURF 56/STRIPE GRN 20, Rx, Sterile; 00597204123 Articular Surface CR ART SURF 56/STRIPE GRN 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Reason
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Identifiers
| code_info | all codes |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2032-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.