RecallRadar

FDA Device recall Z-2031-2025

Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2…

On 2025-07-16 the FDA classified a Class I recall of Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2… by Baxter Healthcare, citing the Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury fr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2031-2025
ClassificationClass I
Statusongoing
Initiated2025-05-30
Published2025-07-16
Last updated2026-09-13 11:57 UTC
CompanyBaxter Healthcare
Distributionn/a

Product

Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. .

Reason

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Identifiers

code_info
All Q-link 13 manufactured between 8/27/2013 to 2/27/2025: 1) Product Code 2010004, UDI/DI 00887761981775; 2) Produc…All Q-link 13 manufactured between 8/27/2013 to 2/27/2025: 1) Product Code 2010004, UDI/DI 00887761981775; 2) Product Code 2040044, UDI/DI 00887761981867; 3) Product Code 2040043, UDI/DI 00887761981850; 4) Product Code 2040045A, UDI/DI 00887761943193; 5) Product Code 2040006, UDI/DI N/A; 6) Product Code 2040007, UDI/DI N/A; 7) Product Code 2030001, UDI/DI N/A. .

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2031-2025%22