FDA Device recall Z-2028-2024

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

On 2024-06-12 the FDA classified a Class II recall of Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual by Remote Diagnostic Technologies, citing simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortion….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2028-2024
ClassificationClass II
Statusongoing
Initiated2024-05-09
Published2024-06-12
Last updated2026-09-13 11:57 UTC
CompanyRemote Diagnostic Technologies
DistributionUS

Product

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Reason

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Identifiers

code_info00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior