FDA Device recall Z-2027-2014
The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to fa…
- FDA Device
- Class II
- terminated
- Published 2014-07-16
On 2014-07-16 the FDA classified a Class II recall of The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to fa… by Hospira, citing hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41,….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2027-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-06-06 |
| Published | 2014-07-16 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hospira |
| Distribution | US |
Product
The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physicians orders.
Reason
Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a
Identifiers
| code_info | 1) Software Version 5.5, List Number: 16037-04-13; 2) Software Version 5.8.1, List Number: 16037-04-31; 3) Software Version 5.8.2, List Number: 16037-04-33 |
|---|