FDA Device recall Z-2026-2019

AXIOM Artis TC, Model Number 7728350

On 2019-07-31 the FDA classified a Class II recall of AXIOM Artis TC, Model Number 7728350 by Siemens Medical Solutions Usa, citing A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the r….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2026-2019
ClassificationClass II
Statusterminated
Initiated2019-06-12
Published2019-07-31
Last updated2026-09-13 11:55 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

AXIOM Artis TC, Model Number 7728350

Reason

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Identifiers

code_infoSerial Numbers: 30006

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.