FDA Device recall Z-2026-2014

The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general s…

On 2014-07-16 the FDA classified a Class II recall of The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general s… by Steris, citing STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor loc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2026-2014
ClassificationClass II
Statusterminated
Initiated2014-06-27
Published2014-07-16
Last updated2026-09-13 11:53 UTC
CompanySteris
DistributionUS

Product

The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.

Reason

STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.

Identifiers

code_infoAll Model Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.