FDA Device recall Z-2025-2019

AXIOM Artis dFC, Model Number 7727717

On 2019-07-31 the FDA classified a Class II recall of AXIOM Artis dFC, Model Number 7727717 by Siemens Medical Solutions Usa, citing A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the r….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2025-2019
ClassificationClass II
Statusterminated
Initiated2019-06-12
Published2019-07-31
Last updated2026-09-13 11:55 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

AXIOM Artis dFC, Model Number 7727717

Reason

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Identifiers

code_infoSerial Numbers: 40072 40031 40075 40070 40032 40067 40064 40065 40061 40017 40066 40071 40059 40038 40029 40039 40048 40073 40062