RecallRadar

FDA Device recall Z-2025-2012

OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or…

On 2012-07-25 the FDA classified a Class II recall of OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or… by Ge Oec Medical Systems, citing please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imag….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2025-2012
ClassificationClass II
Statusterminated
Initiated2007-11-19
Published2012-07-25
Last updated2026-09-13 11:53 UTC
CompanyGe Oec Medical Systems
Distributionn/a

Product

OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or percutaneous surgical procedures. The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgery technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.

Reason

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Identifiers

code_info
Part numbers: 00-887200-01; 00-887200-03; 00-887201-01; 00-887201-03; 00-887202-01; 00-887202-03; 00-887204-03; 00-8872…Part numbers: 00-887200-01; 00-887200-03; 00-887201-01; 00-887201-03; 00-887202-01; 00-887202-03; 00-887204-03; 00-887207-03; 00-887206-03. Along with all associated serial numbers.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2025-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.