FDA Device recall Z-2024-2019

AXIOM Artis dFA, Model Number 7555373

On 2019-07-31 the FDA classified a Class II recall of AXIOM Artis dFA, Model Number 7555373 by Siemens Medical Solutions Usa, citing A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the r….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2024-2019
ClassificationClass II
Statusterminated
Initiated2019-06-12
Published2019-07-31
Last updated2026-09-13 11:55 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

AXIOM Artis dFA, Model Number 7555373

Reason

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Identifiers

code_info
Serial Numbers: 50063 50059 50006 50064 50189 50044 50074 50101 51719 50062 50065 50087 50167 50039 50084…Serial Numbers: 50063 50059 50006 50064 50189 50044 50074 50101 51719 50062 50065 50087 50167 50039 50084 51706 51707 50048 50161 50093 50206 50057 50081 50133 50040 50016 50120 50191 51701 50027 51700 50139 50004 50010 50024 50142 50079 50132 50050

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2024-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.