FDA Device recall Z-2022-2014
STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches
- FDA Device
- Class II
- terminated
- Published 2014-07-16
On 2014-07-16 the FDA classified a Class II recall of STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches by B Braun Medical, citing complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2022-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-29 |
| Published | 2014-07-16 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | B Braun Medical |
| Distribution | US |
Product
STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches
Reason
Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.
Identifiers
| code_info | Product codes: 473547 and 473549 lot #s and expiration dates: 61356774, 1/31/2019; 61362114, 2/28/2019; and 61367665, 3/31/2019. |
|---|