FDA Device recall Z-2016-2026
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS8…
- FDA Device
- Class II
- ongoing
- Published 2026-05-06
On 2026-05-06 the FDA classified a Class II recall of Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS8… by Medline Industries, citing during an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2016-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-03-19 |
| Published | 2026-05-06 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | n/a |
Product
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS840402AI; 2. OPEN HEART CDS, Medline Kit Number/SKU CDS984890V; 3. RR-NEURO VASCULAR PACK-LF, Medline Kit Number/SKU DYNJ0394874L; 4. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993M; 5. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933G; 6. FESS-MB PACK-LF, Medline Kit Number/SKU DYNJ0888159I; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861C; 8. POSTERIOR LUMBAR PACK, Medline Kit Number/SKU DYNJ42626B; 9. CV CUTDOWN PACK, Medline Kit Number/SKU DYNJ45374F; 10. X RAY CLOSED PACK, Medline Kit Number/SKU DYNJ48949C; 11. RFD- AAA PACK, Medline Kit Number/SKU DYNJ54242B; 12. NEURO LAM, Medline Kit Number/SKU DYNJ64188A; 13. LIVE OAK SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ69223; 14. FEM POP PACK, Medline Kit Number/SKU DYNJ81423C; 15. ENS FREEFLAP PROCDURE - LF, Medline Kit Number/SKU DYNJ88249; 16. OPEN HEART A&B, Medline Kit Number/SKU DYNJ901203D; 17. NEURO ACD, Medline Kit Number/SKU DYNJ904549F; 18. KIT VP SHUNT RFD, Medline Kit Number/SKU DYNJ906303F; 19. ACDF-LF, Medline Kit Number/SKU DYNJ906681C; 20. ACDF-LF, Medline Kit Number/SKU DYNJ906681D; 21. ACDF-LF, Medline Kit Number/SKU DYNJ906681F; 22. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196B; 23. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196C; 24. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ908362F; 25. VP SHUNT, Medline Kit Number/SKU DYNJ909379; 26. VP SHUNT, Medline Kit Number/SKU DYNJ909379A.
Reason
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Identifiers
| code_info | Medline Kit Number/SKU CDS840402AI: UDI/DI each 10195327334024, UDI/DI case 40195327334025, Lot Number: 23JLA668; Med…Medline Kit Number/SKU CDS840402AI: UDI/DI each 10195327334024, UDI/DI case 40195327334025, Lot Number: 23JLA668; Medline Kit Number/SKU CDS840402AI: UDI/DI each 10195327334024, UDI/DI case 40195327334025, Lot Number: 23JLA240; Medline Kit Number/SKU CDS984890V: UDI/DI each 10198459374364, UDI/DI case 40198459374365, Lot Number: 25IMG470; Medline Kit Number/SKU CDS984890V: UDI/DI each 10198459374364, UDI/DI case 40198459374365, Lot Number: 25GMI158; Medline Kit Number/SKU CDS984890V: UDI/DI each 10198459374364, UDI/DI case 40198459374365, Lot Number: 25GMC404; Medline Kit Number/SKU DYNJ0394874L: UDI/DI each 10193489800005, UDI/DI case 40193489800006, Lot Number: 23ILA647; Medline Kit Number/SKU DYNJ0394874L: UDI/DI each 10193489800005, UDI/DI case 40193489800006, Lot Number: 23HLA953; Medline Kit Number/SKU DYNJ0394874L: UDI/DI each 10193489800005, UDI/DI case 40193489800006, Lot Number: 23GLA064; Medline Kit Number/SKU DYNJ0394874L: UDI/DI each 10193489800005, UDI/DI |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2016-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.