RecallRadar

FDA Device recall Z-2015-2012

GE Healthcare, Invasive Blood Pressure Care Cables

On 2012-07-25 the FDA classified a Class II recall of GE Healthcare, Invasive Blood Pressure Care Cables by Ge Healthcare, citing GE Healthcare has become aware through complaints of a potential safety issue with its Invasive Blood Pressure Cable made for use with a Spectramed / BD invasive blood pressure tr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2015-2012
ClassificationClass II
Statusterminated
Initiated2012-06-12
Published2012-07-25
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare
DistributionUS

Product

GE Healthcare, Invasive Blood Pressure Care Cables. Product ID Number: 2016995-001, 2016995-002, 2016995-003, 2016995-004.

Reason

GE Healthcare has become aware through complaints of a potential safety issue with its Invasive Blood Pressure Cable made for use with a Spectramed / BD invasive blood pressure transducer. Due to use of an incorrect O-ring, the mating of the cable with the transducer is faulty. This situation inhibits blood pressure measurement from being monitored correctly.

Identifiers

code_info
Product ID Number Lot Code 2016995-001 xxxx2010E 2016995-002 xxxx2010D 2016995-003 xxxx201…Product ID Number Lot Code 2016995-001 xxxx2010E 2016995-002 xxxx2010D 2016995-003 xxxx2010C 2016995-004 xxxx2010C

Status history