FDA Device recall Z-2013-2020
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90121, UDI # 108888571…
- FDA Device
- Class II
- terminated
- Published 2020-05-27
On 2020-05-27 the FDA classified a Class II recall of K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90121, UDI # 108888571… by K2m, citing the Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2013-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-13 |
| Published | 2020-05-27 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | K2m |
| Distribution | US |
Product
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90121, UDI # 10888857108349, Size 24x4mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
Reason
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Identifiers
| code_info | Lot # FKXB |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2013-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.