RecallRadar

FDA Device recall Z-2013-2013

65597201333 Femur CR POR FEM SURF HDN HATCP CML, Rx, Sterile

On 2013-08-28 the FDA classified a Class II recall of 65597201333 Femur CR POR FEM SURF HDN HATCP CML, Rx, Sterile by Zimmer, citing surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2013-2013
ClassificationClass II
Statusterminated
Initiated2013-06-07
Published2013-08-28
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

65597201333 Femur CR POR FEM SURF HDN HATCP CML, Rx, Sterile; 65597201334 Femur CR POR FEM SURF HDN HATCP CMR, Rx, Sterile; 65597201433 Femur CR POR FEM SURF HDN HATCP DML, Rx, Sterile; 65597201434 Femur CR POR FEM SURF HDN HATCP DMR, Rx, Sterile; 65597201533 Femur CR POR FEM SURF HDN HATCP EML, Rx, Sterile; 65597201534 Femur CR POR FEM SURF HDN HATCP EMR, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Identifiers

code_infoall codes

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2013-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.