FDA Device recall Z-2005-2023
PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P,…
- FDA Device
- Class II
- ongoing
- Published 2023-07-05
On 2023-07-05 the FDA classified a Class II recall of PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P,… by Permobil, citing when multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device cont….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2005-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-05-10 |
| Published | 2023-07-05 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Permobil |
| Distribution | n/a |
Product
PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
Reason
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
Identifiers
| code_info | All units with software versions 1.0.00 to 1.0.04 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.