RecallRadar

FDA Device recall Z-2005-2023

PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P,…

On 2023-07-05 the FDA classified a Class II recall of PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P,… by Permobil, citing when multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device cont….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2005-2023
ClassificationClass II
Statusongoing
Initiated2023-05-10
Published2023-07-05
Last updated2026-09-13 11:56 UTC
CompanyPermobil
Distributionn/a

Product

PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Reason

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Identifiers

code_infoAll units with software versions 1.0.00 to 1.0.04

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.