FDA Device recall Z-2002-2016

Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only

On 2016-06-22 the FDA classified a Class II recall of Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only by Datascope, citing custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2002-2016
ClassificationClass II
Statusterminated
Initiated2016-05-09
Published2016-06-22
Last updated2026-09-13 11:54 UTC
CompanyDatascope
DistributionGA

Product

Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours

Reason

Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.

Identifiers

code_infoLot/batch No: 3000003260, Part number 701050175