RecallRadar

FDA Device recall Z-1997-2026

Heraeus, PALAMIX uno

On 2026-05-06 the FDA classified a Class III recall of Heraeus, PALAMIX uno by Heraeus Medical Dental Division, citing based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1997-2026
ClassificationClass III
Statusongoing
Initiated2026-03-25
Published2026-05-06
Last updated2026-09-13 11:58 UTC
CompanyHeraeus Medical Dental Division
DistributionUS

Product

Heraeus, PALAMIX uno. Material Number: 66057893.

Reason

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Identifiers

code_infoMaterial Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1997-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.