FDA Device recall Z-1997-2013
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for det…
- FDA Device
- Class II
- terminated
- Published 2013-08-28
On 2013-08-28 the FDA classified a Class II recall of Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for det… by Beckman Coulter, citing beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1997-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-07-17 |
| Published | 2013-08-28 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | n/a |
Product
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
Reason
Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
Identifiers
| code_info | Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/3…Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1997-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.