RecallRadar

FDA Device recall Z-1992-2016

Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT…

On 2016-06-22 the FDA classified a Class II recall of Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT… by Med Tec, citing when using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1992-2016
ClassificationClass II
Statusterminated
Initiated2016-03-31
Published2016-06-22
Last updated2026-09-13 11:54 UTC
CompanyMed Tec
Distributionn/a

Product

Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041

Reason

When using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.

Identifiers

code_infoVersions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1992-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.