RecallRadar

FDA Device recall Z-1991-2025

ICU Medical Plum Solos, Item number 400011001

On 2025-06-25 the FDA classified a Class II recall of ICU Medical Plum Solos, Item number 400011001 by Icu Medical, citing internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1991-2025
ClassificationClass II
Statusongoing
Initiated2025-04-22
Published2025-06-25
Last updated2026-09-13 11:57 UTC
CompanyIcu Medical
DistributionUS

Product

ICU Medical Plum Solos, Item number 400011001; infusion pump

Reason

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Identifiers

code_info
UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 600000…UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 6000011

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1991-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.