RecallRadar

FDA Device recall Z-1990-2023

Single Use Repositionable Clip QUICKCLIP PRO 230CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.B

On 2023-06-28 the FDA classified a Class II recall of Single Use Repositionable Clip QUICKCLIP PRO 230CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.B by Olympus, citing complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, pote….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1990-2023
ClassificationClass II
Statusongoing
Initiated2023-05-04
Published2023-06-28
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

Single Use Repositionable Clip QUICKCLIP PRO 230CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.B

Reason

Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

Identifiers

code_infoUDI-DI: 04953170385940 Lot Numbers: 24K, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK