FDA Device recall Z-1990-2020
Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer…
- FDA Device
- Class II
- terminated
- Published 2020-05-20
On 2020-05-20 the FDA classified a Class II recall of Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer… by Covidien, citing temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could le….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1990-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-30 |
| Published | 2020-05-20 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Covidien |
| Distribution | US |
Product
Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Reason
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Identifiers
| code_info | Lot Numbers: 1826714264, 1908533364, 1909302964, 1909303064, 1910654864, 1911300664, 1916240464, 1921106564, 1922533864, 2004908564, 2005617264, 2006339564 |
|---|