RecallRadar

FDA Device recall Z-1989-2020

Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Tempe…

On 2020-05-20 the FDA classified a Class II recall of Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Tempe… by Covidien, citing temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could le….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1989-2020
ClassificationClass II
Statusterminated
Initiated2020-04-30
Published2020-05-20
Last updated2026-09-13 11:55 UTC
CompanyCovidien
DistributionUS

Product

Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Reason

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Identifiers

code_infoLot Numbers: 1909308664, 1909320964 1933802964

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1989-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.