FDA Device recall Z-1989-2020
Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Tempe…
- FDA Device
- Class II
- terminated
- Published 2020-05-20
On 2020-05-20 the FDA classified a Class II recall of Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Tempe… by Covidien, citing temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could le….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1989-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-30 |
| Published | 2020-05-20 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Covidien |
| Distribution | US |
Product
Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Reason
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Identifiers
| code_info | Lot Numbers: 1909308664, 1909320964 1933802964 |
|---|