FDA Device recall Z-1987-2021
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT
- FDA Device
- Class II
- ongoing
- Published 2021-07-07
On 2021-07-07 the FDA classified a Class II recall of Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT by Philips North America, citing significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in pot….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1987-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-05-01 |
| Published | 2021-07-07 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT
Reason
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Identifiers
| code_info | Model/ Serial Number/UDI: N/A System Code System Serial Number Unique Device Identifier (UDI#) 882442 2155 (01)008848…Model/ Serial Number/UDI: N/A System Code System Serial Number Unique Device Identifier (UDI#) 882442 2155 (01)00884838059443(21)2155 882442 2154 (01)00884838059443(21)2154 882442 2166 (01)00884838059443(21)2166 882442 2171 (01)00884838059443(21)2171 882442 2153 (01)00884838059443(21)2153 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1987-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.