FDA Device recall Z-1986-2021
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT
- FDA Device
- Class II
- ongoing
- Published 2021-07-07
On 2021-07-07 the FDA classified a Class II recall of Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT by Philips North America, citing significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in pot….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1986-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-05-01 |
| Published | 2021-07-07 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Reason
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Identifiers
| code_info | Model/ Serial Number/UDI: N/A 10016 (01)00884838101111(21)10016 10018 (01)00884838101111(21)10018 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1986-2021%22