RecallRadar

FDA Device recall Z-1983-2019

Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model

On 2019-07-24 the FDA classified a Class II recall of Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model by Olympus, citing WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1983-2019
ClassificationClass II
Statusterminated
Initiated2019-06-10
Published2019-07-24
Last updated2026-09-13 11:55 UTC
CompanyOlympus
DistributionUS

Product

Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664

Reason

WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds.