FDA Device recall Z-1983-2014
Anspach Pneumatic and Electric Motor Systems
- FDA Device
- Class II
- terminated
- Published 2014-07-09
On 2014-07-09 the FDA classified a Class II recall of Anspach Pneumatic and Electric Motor Systems by The Anspach Effort, citing several attachments and handpieces failed to meet the expected sterility assurance level (SAL) when sterilized by the method that is specified in the current Directions for Use (D….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1983-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-08 |
| Published | 2014-07-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | The Anspach Effort |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
Product
Anspach Pneumatic and Electric Motor Systems. Intended for cutting and shaping bone, including the spine and cranium.
Reason
Several attachments and handpieces failed to meet the expected sterility assurance level (SAL) when sterilized by the method that is specified in the current Directions for Use (DFU) provided with the device.
Identifiers
| code_info | DFU: 18-0047: Revisions A & B 18-0074: Revision A; and 18-0105: Revisions A. Operating Manuals: 19-0010: R…DFU: 18-0047: Revisions A & B 18-0074: Revision A; and 18-0105: Revisions A. Operating Manuals: 19-0010: Revisions A, and 19-0013: Revisions A, B, & C. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1983-2014%22