RecallRadar

FDA Device recall Z-1981-2026

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP AN…

On 2026-05-13 the FDA classified a Class I recall of Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP AN… by American Contract Systems, citing impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1981-2026
ClassificationClass I
Statusongoing
Initiated2026-03-19
Published2026-05-13
Last updated2026-09-13 11:58 UTC
CompanyAmerican Contract Systems
DistributionMN

Product

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Reason

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Identifiers

code_info
ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-833…ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1981-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.