RecallRadar

FDA Device recall Z-1981-2025

UROSKOP Omnia

On 2025-06-25 the FDA classified a Class II recall of UROSKOP Omnia by Siemens Medical Solutions Usa, citing A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1981-2025
ClassificationClass II
Statusongoing
Initiated2025-06-04
Published2025-06-25
Last updated2026-09-13 11:57 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

UROSKOP Omnia. Model Number: 10094910

Reason

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Identifiers

code_info
Model Number: 10094910. No UDI. Serial Numbers: 3054, 3029, 3062, 3169, 3402, 3103, 3053, 3055, 3072, 3019, 3048, 3403,…Model Number: 10094910. No UDI. Serial Numbers: 3054, 3029, 3062, 3169, 3402, 3103, 3053, 3055, 3072, 3019, 3048, 3403, 3058, 1045, 2014, 3091, 3013, 3023, 3115, 3051, 3417, 3160, 2033, 3163, 3052, 3166, 3099, 3138, 3157, 3038, 3131, 3037, 3117, 3118, 3096, 3123, 3124, 2022, 2032, 2031, 3134, 3106, 3415, 3042, 3014, 3012, 3108, 3105, 3057, 2008, 3122, 3034, 3151, 3414, 2030, 3021, 3119, 3086, 3073, 3089, 3036, 3031, 3409, 3064, 3153, 3095, 3145, 3079, 3017, 3025, 3076, 2046, 2047, 2055, 2054, 2049, 3416, 2015, 1041, 3155, 3418, 3081, 3084, 2045, 3015, 3125, 3132, 3143, 3085, 3405, 2041, 2009, 3410, 2029, 3404.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1981-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.