RecallRadar

FDA Device recall Z-1981-2012

Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile

On 2012-07-18 the FDA classified a Class II recall of Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile by Zimmer, citing zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated duri….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1981-2012
ClassificationClass II
Statusterminated
Initiated2012-06-05
Published2012-07-18
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: PathFinder NXT is a minimally invasive pedicle screw system. The PathFinder NXT System consists of various screws and rods intended to provide temporary stabilization following surgery to fuse the spine. The intended use of the Fixed Angled Percutaneous Rod Inserter is to insert a rod percutaneously through the extender sieeves to position into the implanted screw heads for securing with closure tops forming a construct that provides temporary stabilization following MIS or mini-open surgery to fuse the spine.

Reason

Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.

Identifiers

code_infoPart numbers 3573-2 (straight handle) all lots.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1981-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.