RecallRadar

FDA Device recall Z-1980-2025

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

On 2025-06-25 the FDA classified a Class II recall of GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S by Ohio Medical, citing erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purit….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1980-2025
ClassificationClass II
Statusongoing
Initiated2025-05-08
Published2025-06-25
Last updated2026-09-13 11:57 UTC
CompanyOhio Medical
DistributionUS

Product

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Reason

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

Identifiers

code_info
Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL…Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837.