FDA Device recall Z-1972-2023

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System…

On 2023-06-21 the FDA classified a Class II recall of Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System… by Boston Scientific Neuromodulation, citing pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, whic….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1972-2023
ClassificationClass II
Statusongoing
Initiated2023-05-12
Published2023-06-21
Last updated2026-09-13 11:56 UTC
CompanyBoston Scientific Neuromodulation
DistributionUS

Product

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

Reason

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Identifiers

code_info
Driver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Supe…Driver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Superion IDS Kit IFU (92479820-02), Surgical Technique Manual (92479821-02)