FDA Device recall Z-1972-2013
Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hem…
- FDA Device
- Class II
- terminated
- Published 2013-08-21
On 2013-08-21 the FDA classified a Class II recall of Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hem… by Edwards Lifesciences, citing edwards Lifesciences is recalling certain lots of the EndoReturn Arterial Cannula that were built with an incorrect component. This component can create a performance problem with….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1972-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-07-16 |
| Published | 2013-08-21 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Edwards Lifesciences |
| Distribution | US |
Product
Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hemostasis valve), an introducer, a guidewire,and a connector hub. The ThruPort catheter introducer sheath kit includes a wire reinforced cannula with hemostasis valve, an introducer, and a guidewire. The cannulae have a wire reinforced section to provide kink resistance and flexibility, tapered tips to aid in insertion and advancement into the femoral artery, a non-reinforced clamp site, a 3/8 in. (9.5 mm) barbed connector, and a hemostasis valve (available on some versions) that allows passage of catheters, such as the EndoClamp aortic catheter. The introducers accept a .038 in. (0.97 mm) guidewire and are marked to simplify assembly and indicate alignment. On some versions, a connector hub secures and immobilizes the introducer within the cannula for easier, one person insertion of the cannula/introducer assembly.
Reason
Edwards Lifesciences is recalling certain lots of the EndoReturn Arterial Cannula that were built with an incorrect component. This component can create a performance problem with the IntraClude IntraAortic Occlusion device.
Identifiers
| code_info | Model ER21B Lots: 59496984, 59516933, 59516936, 59531385, 59531386 , 59543708 Model ER23B Lots: 59464327, 59484349,…Model ER21B Lots: 59496984, 59516933, 59516936, 59531385, 59531386 , 59543708 Model ER23B Lots: 59464327, 59484349, 59484350, 59484351, 59512146, 59513272, 59516940, 59534430, 59535739, 59541763 |
|---|