epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium…

On 2013-08-21 the FDA classified a Class II recall of epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium… by Epocal, citing this recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration valu….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1971-2013
ClassificationClass II
Statusterminated
Initiated2013-03-13
Published2013-08-21
Last updated2026-09-13 11:53 UTC
CompanyEpocal
DistributionUS

Product

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.

Reason

This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.

Identifiers

code_info
LOT #'s EXP. DATES 07-12283-00 03/26/2013 07-12284-00 03/27/2013 07-12285-00 03/28/20…LOT #'s EXP. DATES 07-12283-00 03/26/2013 07-12284-00 03/27/2013 07-12285-00 03/28/2013 07-12289-00 04/01/2013 07-12290-00 04/02/2013 07-12292-00 04/04/2013 07-12293-00 04/05/2013 07-12296-00 04/08/2013 07-12297-00 04/09/2013 07-12298-00 04/10/2013 07-12300-00 04/12/2013 07-12303-00 04/15/2013 07-12305-00 04/17/2013 07-12306-00 04/18/2013 07-12307-00 04/19/2013 07-12310-00 04/22/2013 07-12312-00 04/24/2013 07-12313-00 04/25/2013 07-12314-00 04/26/2013 07-12317-00 04/29/2013 07-12320-00 05/02/2013 07-12321-00 05/03/2013 07-12324-00 05/06/2013 07-12325-00 05/07/2013 07-12327-00 05/09/2013 07-12331-00 05/13/2013 07-12333-00 05/15/2013 07-12335-00 05/17/2013 07-12338-00 05/20/2013 07-12340-00 05/22/2013 07-12341-00 05/23/2013 07-12345-00 05/27/2013 07-1234