Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use…

On 2013-08-21 the FDA classified a Class II recall of Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use… by Terumo Cardiovascular Systems, citing the operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1969-2013
ClassificationClass II
Statusterminated
Initiated2013-07-26
Published2013-08-21
Last updated2026-09-13 11:53 UTC
CompanyTerumo Cardiovascular Systems
DistributionUS

Product

Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.

Reason

The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

Identifiers

code_infoCatalog number: 801764

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.