FDA Device recall Z-1967-2019

BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975 Product Usage: The Microtainer¿ Brand Tube wi…

On 2019-07-24 the FDA classified a Class II recall of BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975 Product Usage: The Microtainer¿ Brand Tube wi… by Becton Dickinson And, citing the products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or micro clots that are not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1967-2019
ClassificationClass II
Statusterminated
Initiated2019-06-03
Published2019-07-24
Last updated2026-09-13 11:55 UTC
CompanyBecton Dickinson And
DistributionAL, CA, CT, FL, GA, IL, KY, MD, MN, MO, OH, PA, PR, TN, TX, VA

Product

BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975 Product Usage: The Microtainer¿ Brand Tube with EDTA(K2) Reorder No. 365974, is used to collect, anticoagulate, transport and store skin puncture blood specimens for testing the following hematological parameters: WBC, RBC ,HGB, HCT, MCV, MCH, MCHC, Platelets, 5 part differential including : Neutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils, Manual WBC Differentials, Reticulocytes, and Spun Hematocrit.

Reason

The products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or micro clots that are not easily detected during visual inspection of the tubes. A clotted sample may lead to recollection of samples or, retesting of patients, resulting in delayed reporting of test results and patient treatment.

Identifiers

code_infoLot Numbers: 9017588 9017590 9017591 9017592 9017593 9017596 9017600 9028803 9045873 9046822 9050660 9052823 9056896

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.