FDA Device recall Z-1966-2025

Allon 2001

On 2025-06-18 the FDA classified a Class II recall of Allon 2001 by Belmont Instrument, citing an investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconn….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1966-2025
ClassificationClass II
Statusongoing
Initiated2025-05-13
Published2025-06-18
Last updated2026-09-13 11:57 UTC
CompanyBelmont Instrument
DistributionUS

Product

Allon 2001. Part Number: 200-00263

Reason

An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

Identifiers

code_info
Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 999390…Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 9993901600001, 9993901600002, 9993901600004, 9993901600006, 9993901600007, 9993901600009, 9995356300005, 9995356300006, 9995356300007, 9995356300008, 9996673800002, 99111143000001, 99111143000003, 99111143000004, 99111143000007, 99111143000008, 99111143000009, 99111143000011, 99111143000012, 99111143000013, 99111143000015, 99111143000018, 99111143000019, 99111288700016, 99111288700021, 99111288700035, 99111288700036, 99111627300011, 99111627300017, 99111627300022, 99111627300029, 99111627300040, 99111627300041, 99111627300045, 99111627300046, 99111627300054, 99111627300057, 99111627300064, 99111627300067, 99113168300004, 99113168300005, 99113168300008, 99113168300010, 99113168300012, 99128623400008, 99129614700002, 99129614700005, 99131046400001, 99131046400003, 99131046400004, 99131046400005, 99131046400006, 99131046400009, 99131046400011, 99131046500007, 99132732400009, 99

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1966-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.