RecallRadar

FDA Device recall Z-1965-2016

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determin…

On 2016-06-22 the FDA classified a Class II recall of TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determin… by Integra Lifesciences, citing as a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Dept….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1965-2016
ClassificationClass II
Statusterminated
Initiated2016-05-10
Published2016-06-22
Last updated2026-09-13 11:54 UTC
CompanyIntegra Lifesciences
DistributionCO, FL, IL, MI, NJ, NY, SC, TX, VA

Product

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

Reason

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Identifiers

code_infoLot UP15H2, Lot UP09D3, Lot UP07E2

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.