Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500

On 2015-07-08 the FDA classified a Class II recall of Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500 by Hamilton Medical, citing an issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator m….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1964-2015
ClassificationClass II
Statusterminated
Initiated2015-06-01
Published2015-07-08
Last updated2026-09-13 11:53 UTC
CompanyHamilton Medical
DistributionUS

Product

Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.

Reason

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.

Identifiers

code_infoPN 155500, all lot codes.