RecallRadar

FDA Device recall Z-1962-2025

2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600

On 2025-06-18 the FDA classified a Class II recall of 2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600 by Baxter Healthcare, citing the replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to e….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1962-2025
ClassificationClass II
Statusongoing
Initiated2025-05-09
Published2025-06-18
Last updated2026-09-13 11:57 UTC
CompanyBaxter Healthcare
DistributionUS

Product

2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600

Reason

The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.

Identifiers

code_infoUDI/DI 00732094139488, Lot Number 25-043, Exp. August 2027

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1962-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.