FDA Device recall Z-1960-2014
ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581)…
- FDA Device
- Class II
- terminated
- Published 2014-07-09
On 2014-07-09 the FDA classified a Class II recall of ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581)… by Siemens Healthcare Diagnostics, citing ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1960-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-06-25 |
| Published | 2014-07-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581), SMN 10378677. Used on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP platforms. For in vitro diagnostic use in the quantitative determination of total homocysteine (HCY) in serum or plasma.
Reason
ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (IFU).
Identifiers
| code_info | Lot #/Exp. Date: 88D1524 01 Jun 2014 88D3303 26 Nov 2013 |
|---|
Status history