RecallRadar

FDA Device recall Z-1958-2014

Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct

On 2014-07-09 the FDA classified a Class II recall of Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct by Focus Diagnostics, citing focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer comp….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1958-2014
ClassificationClass II
Statusterminated
Initiated2014-05-15
Published2014-07-09
Last updated2026-09-13 11:53 UTC
CompanyFocus Diagnostics
DistributionUS

Product

Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413  Simplexa" Flu A/B & RSV Direct. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C.

Reason

Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.

Identifiers

code_info
Model Number: MOL2650. All Lot Numbers. The expiration date of the 614 kits of MOL2650 that are impacted by this Corr…Model Number: MOL2650. All Lot Numbers. The expiration date of the 614 kits of MOL2650 that are impacted by this Correction ranges from 10/31/2014 to 5/31/2015.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1958-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.